Food and drug administration Warns on Insulin Pump Issue

Food and drug administration Warns on Insulin Pump Issue

— Mobile application glitch may well induce t:slim X2 insulin pump battery to drain faster than predicted

by Kristen Monaco , Senior Employees Writer, MedPage Now

A cellular app applied with an insulin pump that led to 224 injuries was recalled by Tandem Diabetic issues Treatment, the Fda introduced right now.

The remember is for the two.7 edition of the Apple iOS t:join mobile application, used in conjunction with t:slender X2 insulin pump with Regulate-IQ know-how, the agency said.

The Food and drug administration determined the motion as a Course I remember, the most significant form. The recall is a correction, not a solution elimination, and was prompted by a software program glitch that may possibly cause the pump battery to drain sooner than predicted. Consumers are currently being urged to update the application to the most up-to-date application.

“Pump shutdown will result in insulin supply to suspend, which could direct to an below-supply of insulin and may perhaps result in hyperglycemia or even diabetic ketoacidosis, which can be a everyday living-threatening problem because of to higher blood sugars and deficiency of insulin,” the Fda warned. As of April 15, no deaths had been documented.

The defective software program may possibly trigger the cell application to crash and be immediately relaunched in the iOS functioning system. This cycle may perhaps intermittently repeat, foremost to extreme Bluetooth interaction that could deplete the pump battery, creating the pump to shut down.

The cell application can be applied to both of those watch pump info and control some areas of the pump through a smartphone. It was launched February 12, 2024, and was offered by March thirteen on the Apple iOS platform.

The pump is intended to deliver insulin subcutaneously for folks with diabetes at established and variable rates. It can connect securely with other equipment like insulin-dosing computer software.

The recall applies to 85,863 gadgets in the U.S.

On March 26, Tandem alerted shoppers and health care vendors working with this process to update to cell app edition 2.7.one or later (accessible in the Apple retailer) and to observe pump battery amount intently. Users ought to charge their system before long immediately after the to start with lower battery notify and carry backup supplies for insulin shipping in scenario the pump fails, the firm stated.

Cleared in December 2019, the t:trim X2 insulin pump was the initial interoperable insulin pump on the marketplace, meaning it can be built-in with many continuous glucose watch sensors. It is offered by prescription only and indicated for use with NovoLog or Humalog U-one hundred insulin.